Radiant Warmer
Add To Cart Download

56-0110

HEATER ASSEMBLY
The heater assembly located above the centre of the support structure consists of an LED examination light with dimming function and a 650 watts heating system (Silica/ quartz /Ceramic Heater) for quick warming and even distribution of heat on the baby bed.
The heater can be swivelled in either direction for taking X-ray and doing other procedures.

AIR MODE
Comes with Air Sensor, added security cover with exclusive feature of switching the heater in case the air temperature goes above 39 ºC irrespective of what the skin temperature is at the time of event.

LED PHOTOTHERAPY OPTIONAL
Comes with highly efficient and bright 12/15 LED on each side of heater. The strong Lux of 24/30 LEDs cure jaundice in baby rapidly.

MANUAL MODE
The system automatically start heating from 40% in manual mode and same can be changed through setting up to 100%.

The system automatically converts from servo mode to manual mode in case of probe failure thus providing nonstop working of the unit.

THE FIRST BREATH
The open care system combine virtually all the technology a caregiver may need in situation needed for resuscitation. The neonate can be supplied with a constant oxygen flow through the oxygen humidifier when connected to an oxygen cylinder/wall unit. The motorized slow suction can be used with a simple flip of a switch. A digital APGAR timer is standard, which signals when 1,5 and 10 minutes have elapsed.

STANDARD COMPOSITION

  • Temperature Control Mode: Pre warm, Automatic, Servo control & manual control;
  • Advanced microprocessor control system with self check on the electronic circuitry skin/air/manual
  • Skin temperature  display can be converted from ºC to ºF
  • Instant Parameter Display: set temp, skin temp, heat power rate;
  • Built-in APGAR Timer and >37°Cpreset function
  • Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;
  • Tilting Bed with Rotary Radiant Head;
  • X-ray Cassette Tray under infant bed, examination light of high luminance;
  • Use anti-blast quartz infrared tube as radiant heating source.
  • Built in SpO2 module to measure O2 saturation and pulse rate.
  • Integrated T-Piece resuscitator with adjustable control of PIP and PEEP
  • Built Suction Unit controlled from main panel
  • Integrated CPAP with built in blender for controlling levels of Oxygen concentration from 21% to 100% with medical air compressor.
  • Integrated Timer : 1 to 59 minute with count - up and count-down features.
  • Storage: The unit comes fitted with slides out for keeping consumables
  • Space for Monitor or accessories.
Technical Specifications
Power AC 110 / 220 V, 50Hz / 60Hz
Input Power ≤1000VA
Heated System Silica/Quartz/Ceramic Heater 650 watts
Phototherapy Heater is also equipped with highly efficient and bright 12/15 LED on both side of heater to
provide light intensity effective on the baby bed
#span#
Fuse 5 Amp
Timer Count Up/ Down
Parameter Display Set temp, skin temp, Air temp. heat power rate, SpO2, Pulse Rate, A big display is
provided for more information and visualization in servo and manual mode
#span#
Control mode Pre-warm mode, auto, manual mode, Servo mode
Temperature
control range
and relative
specification
Temperature control range 25°C ~ 37°C, for set temperature ≤37°C
37°C ~ 38°C, for set temperature >37°C
Skin temperature sensor measure range 25°C~40°C
Air temperature sensor measure range 25°C~42°C
Deviation between sensor measure
temperature and control temperature
≤0.5°C
Skin Sensor Precision ≤0.3°C
Bed surface temperature Uniformity ≤1°C
Temperature Rise Time ≤45min
Baby bed Inclination Tiltable ±12 deg
Alarm Sensor failure alarm Alarm with audible and visual when sensor open circuit,short circuit or disconnect, cut off power
Temperature alarm Alarm with audible and visual when displayed
temperature approach ≥40°C ( or above set temperature)
Deviation alarm After temperature equilibrium, alarm with audible & visual when displayed temperature was 1°C above or below control temperature, cut off power if above 1°C
Power failure alarm Alarm with audible and visual when power failure
APGAR time Alarm with audible when run between 50
seconds to 1 minute, 4 minutes 50 seconds
to 5 minutes, 9 minutes 50 seconds to 10 minutes
Environment Temp.
(It is a suggestion
that please do not
use outside of the
allow environment.)
Operate Temperature Range 18°C~30°C
Store Temperature Range -20°C~+55°C
Environment
Humidity
Operate humidity range 30%RH~75%RH
Store humidity range ≤93%RH
Atmospheric
Pressure
Transport and Store atmospheric
Pressure range
500hPa~1060hPa
Operate atmospheric pressure range 700hPa~1060hPa
Stream velocity Ambient environment stream velocity ≤0.3m/s
Mattress Temp.
Uniformity
≤2.0°C
Temp. rise time ≤45 min
Heat output indicate 0~100%
Radiant box horizon
Angle adjust range
0~±90°step-less
Mattress Tray Size 68cm×56cm (Approx)
Mattress tilt
adjust range
0~±12° step-less
Mattress Tray
Max. load
10 kg (Approx)
Transfusion pole
Max. load
2 kg
Shelf Max. load 2 kg
X-ray Yes
APGAR Timer Yes
Temp. deviation alarm ±1.0°C
Over temp. alarm >40°C ( or above set temperature)
Pressure adjust range Yes
Operating noise Yes
IV Pole Yes
Space for Monitor Yes
Light Source High Bright Blue/ White LED Wavelength – 420- 480 μm
Irradiance <35μm
Storage The unit comes fitted with slides out for keeping consumables
Caster The system is fitted with 4 high quality casters having brakes
Accessories Castor :The system is tted with 4 high quality castors having brakes in front.
Skin Probe, Air Probe, IV Pole , Memory Foam Mattress, Acrylic Helmet, Eye Pads