Enquire Now

Enquire Now

Fill the details to complete enquire now

Closed Incubator
Add To Cart Download

56-0125

Introduction

Infant Incubator is one of the important equipment used in medical institutions. It is mainly used for constant temperature cultivation, temperature resuscitation, oxygen delivery, rescue, hospitalization and observation of sick infants and premature infants. It is used in hospitals, confinement centers, neonatology etc.

 

Features

  • LCD display.
  • Double control modes of bed temperature and skin temperature.
  • Double-layer thermostatic cover, automatic air curtain device.
  • Humidity control function.
  • Air&Oxygen blender.
  • Oxygen supply system.
  • Baby cot is infinitely adjustable.
  • 24 kinds of sound and light alarm, comprehensive protection of baby safety.
General Parameters
  • Power supply : AC220V, 50/60Hz(Standard) 1320VA AC110V, 60Hz (Optional)
  • Size(L x W x H) : 950 x 645 x (1250~11450)mm
  • Maximum Heater Output Power (220VAC) : Air chamber heater : 330W, Water tank heater: 155W
  • Auxiliary Network Power Output : AC220V/50Hz, both maximum permissible currents are 220V 1.5A
  • Heater Indication : 0~100%, 10% increasing amount
  • Temperature Control Mode : Bed temperature mode; skin temperature mode
  • Bed Temperature Control Range : 25.0~37.0°C; 37.1~39.0°C (Span mode)
  • Skin Temperature Control Range : 34.0~37.0°C; 37.1~38.0°C (Span mode)
  • Skin Temperature Sensor Display Range : 5.0~65.0°C
  • Heating Time : ≤50min
  • Temperature Variability : ≤0.5°C
  • Temperature uniformity (mattress in horizontal position) : ≤0.8°C
  • Temperature uniformity(mattress in tilted position) : ≤1.0°C
  • Difference Between Average Incubator Temperature And Control Temperature : ≤1.5°C
  • Skin Temperature Sensor Accuracy : ±0.3°C
Humidity Control
  •  The Running Time Of The Tank After It Is Filled With Distilled Water : Humidity below 70%RH, at least 24h
  • Water Tank Capacity : 1500ml
  • Humidity Display Range : 0%RH~99%RH
  • Humidity Control Range : CAH%RH~90%RH (1% per adjustment)
  • Humidity Control Accuracy : ±10%
Oxygen Concentration Control
  • Oxygen Concentration Display Range : 0%~99%
  • Oxygen Concentration Display Resolution : 1%
  • Oxygen Concentration Control Accuracy : ±5% oxygen volume concentration
  • Oxygen Concentration Setting Range : 21%~60% (1% per adjustment)
  • Oxygen Sensor Service Life : The maximum use time is 10000h at 100% oxygen concentration
Transport Storage
  • Temperature : -20℃~+55℃
  • Atmospheric Pressure Range : 500hPa~1060hPa
  • Relative Humidity : ≤93%RH

The Work Environment

  •  Relative Humidity : 30%RH~75%RH
  • Environment Temperature : +20°C~+30°C
  • Pressure Range : 700hPa~1060hPa
  • Service Altitude : ≤2000m
  • Overvoltage Category : II
  • Pollution Level : 2
  • Ambient Air Velocity : ≤0.3m/s
Other Indicators
  • Carbon Dioxide Concentration In The Infant Compartment A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL/min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0.5% after stabilization
  • The Velocity Of Air Flow Over The Baby Mattress : ≤0.35m/s
General Alarm
Power off alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm
System Alarm
Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 fault alarm, analog-to-digital conversion chip 2548 fault alarm, over temperature alarm, temperature heating system failure, humidity heating system failure
Configuration
Model Type I Type II
Temp. Control Bed/Skin Temperature Bed/Skin Temperature
Display LCD LCD
Bed Temp. Control Range 25°C~39°C 25°C~39°C
Skin Temp. Control Range 34°C~38°C 34°C~38°C
Humidity Control Range CAH%~99%RH CAH%~99%RH
Oxygen Control Range 21%~99% /
Cabinet Yes Yes
*Customizable as per requirement.
*T/C